METTLER TOLEDO™ Ion Selective Electrodes (ISE)
A selection of ISE probes for detecting concentration of various free ions.
Marca: METTLER TOLEDO™ 51340400
Questo articolo non è restituibile.
Consulta la politica di reso
Descrizione
- Prior to first use, rinse the electrode with distilled water and soak it in the appropriate standard for 24-48 hours and then fill with reference electrolyte (refer to table above)
- ISA (Ionic strength adjuster) solution required to be added to all calibration standards and samples prior to measurement for accurate results
- Refer to electrode user manual for appropriate ratio of ISA solution to sample volume
- Add filling solution each day before using the electrode (at least 2.5 cm above sample level in beaker)
- Be sure to screw cap tightly immediately after use
- The filling solution inside the electrode should not be allowed to evaporate, as crystallization will result
- For storage between measurements and up to three days, store in the appropriate standard
- For long-term storage, refer to specific ISE electrode user manual
- Sensing membrane will wear over time; each membrane can last about six months with normal use, but actual lifespan will depend on the type of samples that are being measured
- Competing ion interferences are common. Refer to ISE electrode user manual for potential interferences
- Compatible with METTLER TOLEDO SevenExcellence Benchtop, Seven2Go, and SevenGo Portable
- Compatible with InMotion Autosampler to automate measurement and cleaning of up to 303 samples
Specifica
S7-BNC Cable (30281915) | |
12 mm | |
120 mm | |
0°F to 176°F | |
2 to 13 | |
Ion-Selective Electrode | |
1.0 to 1.0-5 mol/L |
0.5 in. | |
4.75 in. | |
Epoxy Shaft | |
Required Solutions: Reference Electrolyte: 1 mol/L KNO3 Ionic Strength Adjuster - 1 mol/L KNO3 | |
perfectION DX235-Cl- Half-Cell Electrode | |
0°C to 80°C |